It is an area percentage
HPLC purity is the area of the main peak divided by the total integrated area at a given wavelength. It is not the mass fraction of peptide in the vial. A 99% pure lot can have 80% peptide content because water and salts are not detected at 214 nm.
It depends on the method
A purity measured with a shallow gradient, a short run or a different wavelength is not directly comparable with another method. Ask for the method and the chromatogram, not only the number.
It does not prove identity
A single peak can be the wrong sequence. Mass spectrometry and, when possible, sequence confirmation are separate checks. Purity and identity are complementary.
How to use it
Use purity to compare the chromatographic profile of lots analyzed with the same method. Use content to calculate the solution. Use mass to confirm identity. No single number does all three.
Spanish guide: qué significa pureza 99.