What you want to know
A stability panel answers one question: does the solution keep its concentration and profile under the conditions of use? It is not a regulatory study; it is an internal control to know how long an aliquot can be used.
Minimum design
- Points: 0, 24, 72 hours and 7 days (adjust to use).
- Conditions: refrigerated, room temperature and, if relevant, light.
- Controls: freshly prepared solution and vehicle.
- Measurement: HPLC or the validated method for the assay.
What to measure
Concentration, secondary peaks, pH change and appearance. If the method cannot distinguish degradation from adsorption loss, add a recovery control.
What to conclude
Do not claim “stable for 7 days” if you measured one point. Report the range, the condition and the variability. If concentration drops below the criterion you defined, shorten the aliquot life.