Nominal volume is not real volume
An uncalibrated pipette, a poorly fitted tip or a viscous solution changes the volume. In a low-concentration assay, a 5% error can be larger than the difference between conditions.
Calibrate
Check the pipette by gravimetry or with a calibration kit. Do it at least once a year and after a drop or a temperature change. Do not assume it “always worked”.
Technique
- Use the correct tip size and fit it properly.
- Aspirate and dispense at the same speed; do not blow out.
- Do not pipette very cold or viscous liquids without correction.
- Change tips between samples.
- Avoid bubbles and foam.
Small volumes
Below 10 µL, relative error grows. Prepare an intermediate dilution and pipette a larger volume, or use a positive-displacement pipette for viscous liquids.
What to record
Pipette, tip, volume, temperature and any incident. If a result is out of control, check pipetting before repeating the assay.