The axes
The x-axis is time in minutes; the y-axis is absorbance, usually at 214 nm for peptides. The main peak is the peptide; smaller peaks are impurities, salts or solvent. The retention time is not an identity test by itself; it depends on the method.
The baseline
A stable baseline and a flat region before and after the peak are signs of a clean run. A drifting baseline, spikes or a noisy trace make integration unreliable. If the baseline is poor, the purity percentage is not comparable with a clean run.
Integration
The software draws a line under the peak and calculates the area. The purity is the main peak area divided by the total area. Small peaks near the main peak can be missed if the integration parameters are too aggressive; ask for the integration report if the purity is close to the specification.
What the chromatogram does not tell you
It does not confirm the sequence, the mass or the peptide content. A single sharp peak can be the wrong molecule. Use the chromatogram with mass spectrometry and content analysis, not instead of them.