Stationary phase
For peptides, reversed-phase C18 is the most common. C8 can help with more hydrophobic molecules; C4 for large proteins. The choice depends on hydrophobicity and size, not a universal rule.
Particle size
3-5 µm particles give good resolution in conventional HPLC; 1.7-2 µm particles are used in UHPLC and require compatible equipment. Smaller particles increase pressure and can improve resolution.
Pore
For small peptides, 100-130 Å is often enough; for proteins, 300 Å prevents exclusion from the pore. If the pore is too small, recovery drops.
Dimensions
A 150 × 4.6 mm column is a starting point for purity methods. Shorter, narrower columns save solvent and time but require adjusting the gradient and injection volume.
What to validate
Resolution, peak symmetry, pressure, retention time and recovery. Do not change the column without revalidating the method: the chromatogram changes.